Quantitative UV-Vis Spectral Deconvolution
Resolve overlapping absorbance bands, estimate concentrations with matrix equations, and judge residual error after fitting.
Theory
For mixtures, total absorbance at each wavelength is the sum of component absorbances. Multiple wavelengths can solve simultaneous concentration equations.
Procedure
- Choose diagnostic wavelengths.
- Apply Beer-Lambert additivity.
- Solve the concentration matrix.
- Inspect residuals for missed components or baseline error.
Observations
- 430 nm dominant A: The mixture has high absorbance at 430 nm where component A is strongest. A430 = epsA430 cA + epsB430 cB. This wavelength strongly constrains component A concentration.
- 610 nm dominant B: Absorbance at 610 nm is driven mainly by component B. Combining 430 and 610 nm gives cA = 0.034 mM and cB = 0.021 mM. Two-wavelength fitting separates the overlapping spectra.
- Baseline drift check: All residuals are positive at the spectrum edges. A constant offset term reduces edge residuals. Baseline correction is needed before reporting concentrations.
- Three-band model: Adding a weak impurity band removes structured residuals near 520 nm. Least-squares fit with three references gives random residuals below 0.005 A. The expanded model is justified by residual analysis.
Interpretation rule
A good deconvolution has chemically reasonable concentrations and small, structureless residuals.
Validation
Residual patterns reveal whether the model is missing a band, has baseline drift, or uses wrong reference spectra.